User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Simpson 260-6XL:Volt-Ohm-Milliammeter DATE: 15-Jul-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 38 NUMBER OF LINES: 169 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R F W 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Lay the Simpson on its back for this calibration. 1.005 DISP Zero the Meter Movement using the Center Screw. 1.006 HEAD {} 1.007 HEAD {* DC VOLTS TEST *} 1.008 HEAD {}DC VOLTS TEST 1.009 DISP Connect the Red Test Lead to the 250mV input terminal 1.009 DISP of the Simpson. Connect the Black Test Lead to the 1.009 DISP COMMON - Terminal of the Simpson. 1.009 DISP 1.009 DISP Connect the other end of the Test Leads to the 1.009 DISP 5700A output terminals. 1.010 DISP Set the Function Selector switch on the left hand side 1.010 DISP of the Test Instrument to + D.C. 1.011 DISP Set the Test Instrument Range to the 2.5V Range. 1.012 JMP 2.001 1.013 EVAL 2.001 5700 250 250mV 2% 2W #! Test Tol 0.005, Sys Tol 2.7e-006, TUR 1851.852 (>= 4.00). 3.001 DISP Move the Red Test Lead from the 250mV input terminal 3.001 DISP to the + input terminal below the Function Selector sw. 3.002 ASK+ R 3.003 5700 1 1.00V 2% 2W #! Test Tol 0.02, Sys Tol 7.2e-006, TUR 2777.778 (>= 4.00). 4.001 5700 2.5 2.50V 2% 2W #! Test Tol 0.05, Sys Tol 1.65e-005, TUR 3030.303 (>= 4.00). 5.001 DISP Set the Test Instrument Function switch to - D.C. 5.002 5700 2.5 -2.50V 2% 2W #! Test Tol 0.05, Sys Tol 1.65e-005, TUR 3030.303 (>= 4.00). 6.001 DISP Set the Test Instrument Function switch to + D.C. 6.002 ASK+ R 6.003 5700 10 1.0V 2/ 2W #! Test Tol 0.2, Sys Tol 7.2e-006, TUR 27777.778 (>= 4.00). 7.001 5700 10 5.0V 2/ 2W #! Test Tol 0.2, Sys Tol 2.9e-005, TUR 6896.552 (>= 4.00). 8.001 5700 10 10.0V 2% 2W #! Test Tol 0.2, Sys Tol 5.4e-005, TUR 3703.704 (>= 4.00). 9.001 5700 25 25.0V 2% 2W #! Test Tol 0.5, Sys Tol 0.00025, TUR 2000.000 (>= 4.00). 10.001 5700 100 100V 2% 2W #! Test Tol 2, Sys Tol 0.0007, TUR 2857.143 (>= 4.00). 11.001 5700 250 250V 2% 2W #! Test Tol 5, Sys Tol 0.0026, TUR 1923.077 (>= 4.00). 12.001 5700 500 500V 2% 2W #! Test Tol 10, Sys Tol 0.0046, TUR 2173.913 (>= 4.00). 13.001 ASK- R 13.002 DISP Move the RED test lead to the 1000V input terminal 13.002 DISP at the lower right of the Simpson. 13.003 5700 1000 1000V 2% 2W #! Test Tol 20, Sys Tol 0.0086, TUR 2325.581 (>= 4.00). 14.001 HEAD {} 14.002 HEAD {* AC VOLTS TEST *} 14.003 HEAD {}AC VOLTS TEST 14.004 DISP Set the Test Instrument Function switch to A.C. 14.005 JMP 15.001 14.006 EVAL 15.001 5700 1000 1000V 3% 400H 2W #! Test Tol 30, Sys Tol 0.084, TUR 357.143 (>= 4.00). 16.001 DISP Move the RED test lead from the 1000V input terminal 16.001 DISP to the + terminal near the bottom left corner of 16.001 DISP the meter. 16.002 5700 500 500V 3% 400H 2W #! Test Tol 15, Sys Tol 0.044, TUR 340.909 (>= 4.00). 17.001 ASK+ R 17.002 5700 250 250V 3% 400H 2W #! Test Tol 7.5, Sys Tol 0.024, TUR 312.500 (>= 4.00). 18.001 5700 100 100V 3% 400H 2W #! Test Tol 3, Sys Tol 0.009, TUR 333.333 (>= 4.00). 19.001 5700 25 25.0V 3% 400H 2W #! Test Tol 0.75, Sys Tol 0.003, TUR 250.000 (>= 4.00). 20.001 5700 10 10.0V 3% 400H 2W #! Test Tol 0.3, Sys Tol 0.00082, TUR 365.854 (>= 4.00). 21.001 5700 2.5 2.500V 3% 400H 2W #! Test Tol 0.075, Sys Tol 0.0002575, TUR 291.262 (>= 4.00). 22.001 HEAD {} 22.002 HEAD {* FREQUENCY RESPONSE TEST *} 22.003 HEAD {}FREQUENCY RESPONSE TEST 22.004 JMP 23.001 22.005 EVAL 23.001 5700 2.5 2.500V 3% 200H 2W #! Test Tol 0.075, Sys Tol 0.0002575, TUR 291.262 (>= 4.00). 24.001 5700 2.5 2.500V 3% 60H 2W #! Test Tol 0.075, Sys Tol 0.0002575, TUR 291.262 (>= 4.00). 25.001 HEAD {} 25.002 HEAD {* DC CURRENT TEST *} 25.003 HEAD {}DC CURRENT TEST 25.004 DISP Set Test Instrument Function switch to + D.C. 25.005 JMP 26.001 25.006 EVAL 26.001 5700 0.5 0.50mA 2% 2W #! Test Tol 1e-005, Sys Tol 3.5e-008, TUR 285.714 (>= 4.00). 27.001 5700 5 5.0mA 2% 2W #! Test Tol 0.0001, Sys Tol 3.5e-007, TUR 285.714 (>= 4.00). 28.001 5700 50 50mA 2% 2W #! Test Tol 0.001, Sys Tol 4e-006, TUR 250.000 (>= 4.00). 29.001 5700 500 500mA 2% 2W #! Test Tol 0.01, Sys Tol 7e-005, TUR 142.857 (>= 4.00). 30.001 ASK- R 30.002 DISP Set the Test Instrument Range to 25V (50u amps). 30.003 DISP Move the Red wire from the Test Instrument + Input 30.003 DISP to the 50 uAMPS Input. 30.004 5700 50 50.00uA 1.5% 2W #! Test Tol 7.5e-007, Sys Tol 1.25e-008, TUR 60.000 (>= 4.00). 31.001 DISP Set the Test Instrument Range to 5mA (5amps). 31.002 DISP Move the Red wire from the meters 50uAMPS 31.002 DISP Input to the +5A Input and the Black lead to the 31.002 DISP -5A Input of the meter. 31.003 ASK+ W 31.004 5700 5 5.0A 2% B1 2W #! Test Tol 0.1, Sys Tol 0.00218, TUR 45.872 (>= 4.00). 32.001 ASK- W 32.002 ASK+ K 32.003 HEAD {} 32.004 HEAD {* RESISTANCE TEST *} 32.005 HEAD {}RESISTANCE TEST 32.006 DISP Move the Red wire from the Test Instruments +5A Input 32.006 DISP to the + Input near the bottom left hand corner. 32.006 DISP Move the Black wire from the -5A Input to the 32.006 DISP - COMMON Input Terminal of the Simpson. 32.007 DISP Set the Range to {Rx1}. 32.008 DISP Short the wires from the Test Instrument together at 32.008 DISP the Test Interface Panel. 32.008 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 32.008 DISP Remove the short. 32.009 ASK+ W 32.010 JMP 33.001 32.011 EVAL 33.001 5700 10.00Z +12.5% -11% 2W #! T.U.R. not calculated because tolerance or floor unspecified. 34.001 DISP Set the Range to {Rx100}. 34.002 DISP Short the wires from the Test Instrument together at 34.002 DISP the Test Interface Panel. 34.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 34.002 DISP Remove the short. 34.003 JMP 35.001 34.004 EVAL 35.001 5700 1.00kZ +8% -7.5% 2W #! T.U.R. not calculated because tolerance or floor unspecified. 36.001 DISP Set the Range to {Rx10,000}. 36.002 DISP Short the wires from the Test Instrument together at 36.002 DISP the Test Interface Panel. 36.002 DISP Adjust the ZERO OHMS control for 0 reading on the meter. 36.002 DISP Remove the short. 36.003 JMP 37.001 36.004 EVAL 37.001 5700 100kZ +8% -7.5% 2W #! Test Tol 7500, Sys Tol 1.3, TUR 5769.231 (>= 4.00). 38.001 HEAD {} 38.002 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 2 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.